Drug Chamber

What is drug chamber?

The drug chamber device is in the form of a chamber that measures the stability of the drug under the influence of the factors mentioned above in the time periods of 3 months, 6 months and 1 year.

After performing the acceleration test with high accuracy and quality, the certificate (Q Validation (PQ, OQ, I) is confirmed.

Drug chamber is an environment that can be controlled in terms of temperature and humidity, which pharmaceutical companies use at a temperature of 30 to 40 and a humidity of 65 to 75%.
The drug chamber is calibrated at this temperature for better performance and is tested for 24 to 48 hours for correct performance.
Also, IQ (Installation Qualification) and OQ (Operation Qualification) tests are performed at the user’s place to ensure the correct operation of the device and confirm it for the user.

models:

1300 liters
600 liters
400 liters
300 liters
150 liters

The expert and technical forces at Noor Sanat Tehzih Ferdous company, benefiting from technical knowledge and knowing the importance of this device in the pharmaceutical and medical industry of the country, design and produce pharmaceutical chambers with high precision and quality, which is one of its most important features. The following can be mentioned:

The outer body is made of oiled sheet with electrostatic paint
The material of the inner body is made of 304 steel
It has a main door and a glass door for the user to visit with 5 cm thick insulation
Equipped with temperature and humidity control
Equipped with circulation fan for uniformity of temperature and humidity
Equipped with audio alarms to warn when necessary
Programmable
Has an information screen
Has a timer
Ability to save information
The possibility of calibrating humidity in 2 points and temperature in 6 points.

Price list and purchase:

989109249050+

Description

What is drug chamber?

Drug chamber is an environment that can be controlled in terms of temperature and humidity, which pharmaceutical companies use at a temperature of 30 to 40 and a humidity of 65 to 75%. The drug chamber is calibrated for better performance at this temperature and is tested for 24 to 48 hours for correct operation, also to ensure the correct operation of the device and confirm it for the user IQ (Installation Qualification) and OQ (Operation Qualification) tests in The place of use is done.

One of the concerns of pharmaceutical users is the performance of the device during use and under load, and to solve this concern, PQ (Necessary cooperating during performance qualification) tests are also performed on site.

Noor Sanat Tehzih Ferdous Company will provide all the mentioned test certificates along with drug chamber device if required by the customer.

Also, to learn more about the chamber device, you can read the article What is a chamber.

Necessity of using drug chamber:

In today’s era, powerful pharmaceutical companies rely on their knowledge and expertise to produce the drugs needed by the country. They have also achieved many successes in this field, which is the result of continuous efforts in the field of production.

One of the most important principles in the research department of pharmaceutical and medical sciences, and especially in the quality control department of pharmaceutical factories, is the focus of the manufacturers’ attention, is to conduct drug stability tests in specific conditions of temperature and humidity and very diverse climates of the country.

Manufacturers are definitely concerned about the impact of external environmental factors such as:

temperature
humidity
The amount of sunlight

In different weather conditions of the country, they are on production drugs.

They believe that drug tolerance should be measured in different conditions. In fact, the lifespan of the drug (Aging Test) should be tested.

Chamber Daro performs a complete analysis using the data logger it has and provides it to the user through a computer connection.

5/5 - (2 votes)
Categories: ,

Technical Specifications

مدلTCH 1300 600 TCH 400 TCH 300 TCH 150 TCH
حجم (lit)1300600400300150
ابعاد خارجی (mm)w:2300 * d:890 * h:1900w:820 * d:970 * h:2000w:750 * d:1070 * h:1870w:750 * d:10700 * h:1635w:700 * d:910 * h:1405
ابعاد داخلی (mm)w:1550 * d:750 * h:1200w:650 * d:725 * h:1150w:560 * d:610 * h:1100w:560 * d:610 * h:870w:520 * d:500 * h:600
جنس بدنه خارجیورق فولادیورق فولادیورق فولادیورق فولادیورق فولادی
جنس بدنه داخلیاستیل 304 استیل 304 استیل 304 استیل 304 استیل 304
طبقه بندی2*44332
ارتفاع طبقاتمتغیرمتغیرمتغیرمتغیرمتغیر
سرعت هوا0.4 ms0.4 ms0.4 ms0.4 ms0.4 ms
محدوده دما 4°c تا 40°c 4°c تا 40°c 4°c تا 40°c 4°c تا 40°c 4°c تا 40°c
یکنواختی دما± 0.5 °c± 0.5 °c± 0.5 °c± 0.5 °c± 0.5 °c
دقت نقطه ای دما± 0.2 °c± 0.2 °c± 0.2 °c± 0.2 °c± 0.2 °c
سنسور دماPT 100PT 100PT 100PT 100PT 100
محدوده رطوبت40% - 95%40% - 95%40% - 95%40% - 95%40% - 95%
یکنواختی رطوبت0.050.050.050.050.05
دقت رطوبت0.030.030.030.030.03
سنسور رطوبتترانسمیترترانسمیترترانسمیترترانسمیترترانسمیتر
سیستم رطوبت سازبخار گرمبخار گرمبخار گرمبخار گرمبخار گرم
کنترلPIDPIDPIDPIDPID
نمایشگردیجیتال دیجیتال دیجیتال دیجیتال دیجیتال
هشدار دهنده صوتیدارددارددارددارددارد
هشدار دهنده مکانیکیدارددارددارددارددارد
حمل و نقلپرتابلپرتابلپرتابلپرتابلپرتابل
سیستم پیامکیدارد دارد دارد دارد دارد
اتصال به شبکهدارددارددارددارددارد
ذخیره اطلاعاتدارد دارد دارد دارد دارد
برودتدانش فنی دانش فنی دانش فنی دانش فنی دانش فنی
برنامه ریزی10 برنامه10 برنامه10 برنامه10 برنامه10 برنامه
عایق بندیفوم تزریقیفوم تزریقیفوم تزریقیفوم تزریقیفوم تزریقی
5/5 - (2 votes)

Download Catalog

چمبر - chamber

5/5 - (2 votes)

more information

Necessary environmental conditions for installing the chamber:

Air temperature: 20 to 30 degrees Celsius
Air humidity: 60% relative humidity
Location: flat and level surface of the device
220V power supply and proper grounding should be used.
It is suggested to use a stabilizer to prevent power fluctuations.
Avoid direct sunlight on the side of the window.